Basic Info.
Model NO.
JIBPT-01-UT
Name
Disposable Blood Pressure Transducer for Utah
Properties
Medical Materials & Accessories
Certificate
CE/ISO13485
Feature
with Flush Valve
Packing
Custom
Sample
Available
Transport Package
Sterile Blister Box
Specification
224*113*45mm
Trademark
JCR
Origin
China
HS Code
90181930
Production Capacity
4000000/Month
Product Description
Disposable IBP Transducer
1.Consistant accurate working during blood pressure monitoring.
2.100000 grade clean workshop and 100% finial test.
3..Sterile and individual pakage.
4 connectors compatible with most monitors.
5.Single/double/triple channel are options.
6. OEM/ODM are available.
7. ISO CE approval
1.Consistant accurate working during blood pressure monitoring.
2.100000 grade clean workshop and 100% finial test.
3..Sterile and individual pakage.
4 connectors compatible with most monitors.
5.Single/double/triple channel are options.
6. OEM/ODM are available.
7. ISO CE approval
Main technical parameters and working environment requirements: | ||||
Disposable Pressure Transducer | ||||
Barometric pressure | 70~106Kpa | |||
Relative humidity | 10~90% (non-condensing) | |||
Operating pressure range | -50 ~ + 300 mmHg | |||
Sensitivity | 5.0μV/V/mmHg ±3% | |||
Nonlinearity and hysteresis | full-scale reading±1.5% | |||
Input impedance | 1200Ω ~3200Ω | |||
Zero pressure offset | -20mmHg~+20mmHg | |||
Thermal offset shift | ≤±0.3mmHg/°C | |||
Offset drift | After warm-up for 20 seconds, drift within 2 mmHg within 8 hours | |||
Thermal span shift | ≤±0.1%/°C | |||
Frequency response | Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz | |||
Defibrillator withstand | Defibrillation depends on the final connection of the equipment. | |||
Leakage current | Leakage current is dependent on the final connection of the equipment | |||
Overpressure load | -400~+4000mmHg | |||
Shock resistance | Withstand falling from one meter for three times | |||
| When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage. | |||
Storage and transport environment requirements | - 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold. | |||
Contact with the human body | ≤168h | |||
Intra-system velocity | ||||
Flow | 2-5mL/h | |||
Intra-system refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions. | ||||
Product shelf-life | 3 years |